JNNP

HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER
[Advanced]

Published Online First: 30 March 2006. doi:10.1136/jnnp.2005.081232
Journal of Neurology, Neurosurgery, and Psychiatry 2006;77:868-872
Copyright © 2006 by the BMJ Publishing Group Ltd.

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
jnnp.2005.081232v1
77/7/868    most recent
Right arrow Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this link to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Add article to my folders
Right arrow Download to citation manager
Right arrowRequest Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Voges, J
Right arrow Articles by Sturm, V
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Voges, J
Right arrow Articles by Sturm, V

PAPER

Deep-brain stimulation: long-term analysis of complications caused by hardware and surgery—experiences from a single centre

J Voges, Y Waerzeggers, M Maarouf, R Lehrke, A Koulousakis, D Lenartz, V Sturm

Department of Stereotaxy and Functional Neurosurgery, University Köln, Köln, Germany

Correspondence to:
Correspondence to:
Professor Dr Jürgen Voges
Department of Stereotaxy and Functional Neurosurgery, University Köln, Kerpenerstr 62, 50924 Köln, Germany; j.voges{at}uni-koeln.de

Objective: To determine the surgery-related and hardware-related complications of deep-brain stimulation (DBS) at a single centre.

Methods: 262 consecutive patients (472 electrodes) operated for DBS in our department from February 1996 to March 2003 were retrospectively analysed to document acute adverse events (30 days postoperatively). The data of 180 of these patients were additionally revised to assess long-term complications (352 electrodes, mean follow-up 36.3 (SD 20.8) months).

Results: The frequency of minor intraoperative complications was 4.2% (11/262 patients). Transient (0.2%) or permanent (0.4%) neurological deficits, and in one case asymptomatic intracranial haemorrhage (0.2%), were registered as acute severe adverse events caused by surgery. Among minor acute complications were subcutaneous bleeding along the extension wire (1.2%) and haematoma at the pulse generator implantation site (1.2%). Skin infection caused by the implanted material was registered in 15 of 262 patients (5.7%). The infection rate during the first observation period was 1.5% (4/262 patients) and the late infection rate was 6.1% (11/180 patients). Partial or complete removal of the stimulation system was necessitated in 12 of 262 (4.6%) patients because of skin infection. During the long-term observation period, hardware-related problems were registered in 25 of 180 (13.9%) patients.

Conclusions: Stereotactic implantation of electrodes for DBS, if performed with multiplanar three-dimensional imaging and advanced treatment planning software, is a safe procedure with no mortality and low morbidity. The main causes for the patients’ prolonged hospital stay and repeated surgery were wound infections and hardware-related complications.


Abbreviations: BMI, body mass index; DBS, deep-brain stimulation; fMRI, functional MRI; ICH, intracranial haemorrhage; IPG, implantable pulse generator; MER, microelectrode recording; OCD, obsessive–compulsive disorders; STN, subthalamic nucleus







HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS REGISTER