J Neurol Neurosurg Psychiatry 2004;75:280-286 doi:10.1136/jnnp.2002.008904
  • Paper

False negative findings in intraoperative SEP monitoring: analysis of 658 consecutive neurosurgical cases and review of published reports

  1. H Wiedemayer1,
  2. I E Sandalcioglu1,
  3. W Armbruster2,
  4. J Regel1,
  5. H Schaefer1,
  6. D Stolke1
  1. 1Department of Neurosurgery, University of Essen, Essen, Germany
  2. 2Department of Anaesthesiology and Intensive Care Medicine, University of Essen
  1. Correspondence to:
 Priv-Doz Dr H Wiedemayer
 Universitätsklinikum Essen, Neurochirurgische Klinik, Hufelandstr 55, 45122 Essen, Germany; helmut.wiedemayeruni-essen.de
  • Received 23 December 2002
  • Accepted 7 May 2003
  • Revised 14 March 2003

Abstract

Objectives: To determine the sensitivity of intraoperative monitoring in neurosurgical operations using somatosensory evoked potentials and to identify reasons for false negative findings and possible settings with an increased risk for monitoring failure.

Methods: SEP monitoring of 658 neurosurgical operations was analysed. The target of monitoring was the function of a hemisphere in 251 cases, the brain stem in 198 cases, and the spinal cord in 209 cases.

Results: In 27 cases (4.1%), monitoring was classified as false negative. Further analysis showed that five of these patients had experienced delayed neurological damage. Among the remaining 22 false negative cases, 14 had a minor neurological deficit and eight had severe neurological damage. Overall sensitivity and negative predictive value of SEP monitoring was 79% and 96%, respectively. For the detection of severe neurological damage the corresponding figures were 91% and 98%. Sensitivity of monitoring varied depending on the target of monitoring and the type of lesion. Monitoring was less likely to detect neurological damage in surgery for infratentorial tumours with brain stem compression, small lesions of the motor cortex, and small vessel damage during aneurysm surgery.

Conclusions: SEP monitoring has acceptable sensitivity for detecting neurological damage during different neurosurgical procedures. Distinct settings with an increased risk of monitoring failure can be identified. In these cases measures to enhance the sensitivity of monitoring should be considered.

Footnotes

  • Competing interests: None declared.

Register for free content

The full back archive is now available for all BMJ Journals. Institutional subscribers may access the entire archive as part of their subscription. Personal subscribers will also have access to all content when logged in. Non-subscribers who register have free access to all articles published before 2006 right back to volume 1 issue 1. Register here to access the free archive of all BMJ Journals.