Table 2

Sensitivity of the Rydel-Seiffer (RS) tuning fork compared with the Vibrameter in patients with a mild polyneuropathy (chosen specificity of 95%)

LocationRS tuning fork: patients with abnormal vibration (n (%))Vibrameter: patients with abnormal vibration (n (%))
Ulnar styloid process17 (29)18 (31)
Index finger19 (32)19 (32)
Internal malleolus34 (58)28 (48)
Hallux45 (76)43 (73)
Sites examined with abnormal sense of vibration:
010 (17)13 (22)
112 (20)10 (17)
219 (32)18 (31)
3 or 418 (31)18 (31)