Comparison of transducers (1=better; 2= worse)

CaminoSpiegelberg
24 hours zero drift11
Temperature drift (27°C-40°C)21
Accuracy in static measurement1(<1 mm Hg/100 mm Hg)2(5 mm Hg/100 mm Hg)
Accuracy in measurement of pulse pressure * Decreasing with mean pressure increasing
Bandwidth * 4 Hz at 10 mm Hg mean pressure
Pulse recording * Peak to peak amplitude underestimated, delayed 0.1s
  • 1-150 These values were measured relative to the Camino monitor.