Assessment | Valve | Risk difference (95% CI) | OR (95% CI)* | p Value | |
---|---|---|---|---|---|
Programmable | Gravitational | ||||
Hospital stay (t1) | |||||
Participants | 71 | 74 | |||
Subdural effusion | 10 | 1 | −13% (−21% to −4%) | 0.08 (0.01 to 0.67) | 0.020 |
Bleeding complications | 4 | 1 | −4% (−10% to 2%) | 0.23 (0.03 to 2.10) | 0.193 |
Misplacement | 6 | 4 | −3% (−11% to 5%) | 0.62 (0.17 to 2.29) | 0.473 |
Follow-up at 6 months (t2) | |||||
Participants | 67 | 70 | |||
Subdural effusion | 24 | 4 | −30% (−43% to −17%) | 0.11 (0.04 to 0.33) | < 0.001 |
Overdrainage | 26 | 4 | −33% (−46% to −20%) | 0.10 (0.03 to 0.29) | < 0.001 |
Cumulative incidence of overdrainage | 29 | 5 | −36% (−49% to −23%) | 0.10 (0.04 to 0.28) | < 0.001 |
Underdrainage | 3 | 2 | −2% (−8% to 5%) | 0.62 (0.10 to 3.82) | 0.604 |
Follow-up at 12 months (t3) | |||||
Participants | 58 | 60 | |||
Subdural effusion | 12 | 0 | −21% (−31% to −10%) | – | |
Overdrainage | 14 | 2 | −21% (−33% to −9%) | 0.10 (0.02 to 0.46) | 0.003 |
Cumulative incidence of overdrainage | 34 | 7 | −44% (−59% to −30%) | 0.10 (0.04 to 0·24) | <0.001 |
Underdrainage | 8 | 1 | −12% (−21% to −3%) | 0.00 (0.01 to 0.84) | 0.033 |
Cumulative incidence of adverse events | |||||
Wound infection | 1 | 2 | 1% (−3% to 6%) | 1.94 (0.17 to 21.93) | 0.591 |
Local bleeding | 4 | 1 | −4% (−10% to 2%) | 0.23 (0.03 to 2.10) | 0.193 |
Malplacement | 6 | 4 | −3% (−11% to 5%) | 0.62 (0.17 to 2.29) | 0.473 |
Device exchange | 3 | 1 | −3% (−8% to 2%) | 0.31 (0.03 to 3.06) | 0.316 |
Pressure adjustment | 29 | 28 | −3% (−19% to 13%) | 0.88 (0.45 to 1.72) | 0.711 |
Death | 0 | 2 | 3% (−1% to 6%) | – |
*Derived from generalised linear model, unadjusted data.