Quantification of fMRI artifact reduction by a novel plaster cast head holder

Hum Brain Mapp. 2000 Nov;11(3):207-13. doi: 10.1002/1097-0193(200011)11:3<207::aid-hbm60>3.0.co;2-j.

Abstract

In light of artifact-induced high variability of activation in fMRI repeat studies, we developed and tested a clinically useful plaster cast head holder (PCH) with improved immobilization, repositioning, and comfort. With PCH, there were considerably lower levels of translational and rotational head motion components compared to head fixation with conventional restraining straps (CRS). Rotational components cannot be fully compensated by realignment and lead to "false activations." In addition, task-correlated head motion, which highly increases the risk of artifacts, was considerably reduced with PCH, especially in a motion prone subject. Compared with PCH, head motion was 133% larger with CRS in a highly cooperative subject. With a motion prone subject, head motion range was increased by 769% (PCH: 0.9 mm, CRS: 7.8 mm), which may indicate the usefulness of PCH for restless patients. In functional activation maps, PCH alone yielded fewer residual motion artifacts than CRS + image registration. Subject tolerance of the head holder during the long measurement times of up to 2.5 hr was good, and slice orientation on different days confirmed the quality of repositioning.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Artifacts*
  • Brain Mapping / methods*
  • Equipment Design
  • Head
  • Humans
  • Immobilization
  • Magnetic Resonance Imaging / methods*
  • Prone Position
  • Reproducibility of Results